he balance should be used carefully.Clean the pan and the case frequently with soft cloth and toothpaste.Don,t wipe the balance with strong agent.
TROUBLESHOOTING
SN | TROUBLE | CAUSE | REMEDY |
1 | No light on Display | ·Power is not on ·Display switch is not on ·Instant interference | ·Check and turn it on again ·Depress ON button ·Switch on again and re-plug power |
2 | Display the “-----” upper line part only | ·Overload ·The calibration in the internal memory may ba damaged ·The pan is not installed correctly | ·Reduce the load immediately ·Recalibrate according to the above procedures. About 8 seconds after the standard weight has been put on,the calibrated result may be displayed. A certain stable time is necessary. ·Take out the pan and reinstall it. |
3 | Display the lower half Only——— | ·without pan,too light ·he pun is not installed correctly | ·Reinstall the pan |
4 | The weighing result is not stable (date changed swiftly) | ·due to air flow | ·Cheek the windproof cover tu see if it is closed. ·Change to a longer integrating time |
5 | The result is not correct | ·not zeroing before weighing ·Use the balance without calibration or the cal weight is not accurate ·The line voltage is too low or net correct | ·Depress TAR button |
SN | TROUBLE | CAUSE | REMEDY |
6 | Display remains at a certain digit or indicates nonsense symbol | ·Instant interference ·Wrong line voltage | ·Turn on the balance once again or re-plug the power cord ·Chang to normal line voltage |
7 | The stable mark “0”on the left of the display does not go out | ·A higher balance sensitivity ·undesirable environment such as strong air flow or vibration | ·Set to lower sensitivity ·Improve the environment |
8 | Remain at Waiting Status———— | ·The balance position is not correct, e.g. With strong air flow, vibration or great fluctuation of room temp. ·The selected sensitivity is too high | Improve the environment ASD-3 |
9 | Cal Err Displays | ·There is an object on the pan before calibration ·not clear before calibration ·Depress CAL before the | ·Take away the object,clear and recalibrate |
10 | Err-1 Err-2 Display | ·Instant interference ·Something wrong with the balance | ·Turn on the balance once again ·Send it to the service center |
11 | The weighing unit on the right dose not display | ·not calibrated ·The calibrated number in the internal memory of the balance has been erased | ·Calibrate the balance |
12 | Cou-Err Display | ·No Constant be set before operating counting function ·Overload when setting constant ·Underload when setting constant | ·Set average number before operating counting function |